Use only
Advocate™
control solutions.
Check the expiration date on the control solution vial.
Do not use if expired.
Control solution, meter, and test strips should come to
room temperature (20-25°C/68-77°F) before testing.
Shake the vial, discard the first drop of control
solution, and wipe off the dispenser tip to ensure a
good sample and an accurate result.
Use only for
90 days after first opening. Record the
discard date (date opened plus 90 days) on the
control solution vial. Discard after 90 days.
Store the control solution tightly closed at temperatures
below 30°C (86°F). Do not refrigerate.
PLEASE NOTE
The control solution range printed on the test strip vial is for
Advocate™
control solution only. It is used to test meter and test strip
performance. It is not recommended range for your blood glucose
Important Control Solution Information