the meter and the test strips are working together as a
system and that you are performing the test correctly. It
is very important that you do this simple check routinely
to make sure you get accurate results.
How often the control solution test should be performed?
When the control solution test should be performed?
When you first get your Glucose Meter.
When you begin using a new vial of test strips.
Whenever you suspect that the meter or test strips are
not working properly.
When your blood glucose test results are not consistent
with how you feel, or when you think your results are
not accurate.
When your test strips are exposed to extreme environmental
conditions.
(See Storage section of this manual).
When you want to practice running the test.
If you drop the meter.
When you use this system to test your blood for the first
time, practice the procedure using control solution. When
you can do three tests in a row that are within the expected
range, you are ready to test your blood.
To routinely check the meter and test strips, perform a
single test for each level of control solution supplied at
least once a week.